Conventional multiple-track-location shear wave elasticity imaging (MTL-SWEI) is a powerful tool for noninvasively estimating tissue elasticity. The resolution and noise levels of MTL-SWEI systems, however, are limited by ultrasound speckle.… Click to show full abstract
Conventional multiple-track-location shear wave elasticity imaging (MTL-SWEI) is a powerful tool for noninvasively estimating tissue elasticity. The resolution and noise levels of MTL-SWEI systems, however, are limited by ultrasound speckle. Single-track-location SWEI (STL-SWEI) is a novel variant which fixes the position of the tracking beam and modulates the push location to effectively cancel out the effects of speckle-induced bias. We present here a 3-D STL-SWEI system, which provides full suppression of lateral and elevation speckle bias for high-resolution volumetric elasticity imaging, and requires no spatial smoothing to make accurate measurements of shear wave speed. We demonstrate and analyze the system’s performance in homogeneous and layered elasticity phantoms.
               