The study of Te, As, Bi, Sb and Se (TABS) has increased over the past years due to their use in the development of low‐carbon energy technologies. However, there is… Click to show full abstract
The study of Te, As, Bi, Sb and Se (TABS) has increased over the past years due to their use in the development of low‐carbon energy technologies. However, there is a scarcity of mass fraction values of TABS in geological reference materials. This underlines the difficulty in undertaking routine determinations of these elements. The mass fractions of TABS were determined in geological reference materials using hydride generation‐atomic fluorescence spectrometry (HG‐AFS), calibrated with standard solutions. Comparisons with literature values were used to validate the method. Samples from the GeoPT proficiency test were also analysed. For most elements, there are no assigned or even provisional values for many of the GeoPT and reference materials because of the wide range of results reported. For mass fractions above the quantification limit of the method, our results are in good agreement with the median of GeoPT results. Thus, we propose GeoPT median values as informational values for these elements. In contrast, at mass fractions < 0.5 µg g−1 median values of Se from GeoPT are systematically higher than our results. Our Se results are in agreement with the reference materials down to 0.02 µg g−1, which suggest that many of the results for Se reported in GeoPT testing are too high.
               