Intrinsic antitachycardia pacing (iATP) is a novel automated ventricular ATP algorithm that designs ATP sequences based on the analysis of prior failed ATP. Real‐world data on the efficacy and safety… Click to show full abstract
Intrinsic antitachycardia pacing (iATP) is a novel automated ventricular ATP algorithm that designs ATP sequences based on the analysis of prior failed ATP. Real‐world data on the efficacy and safety of iATP are lacking. Among 124 ventricular tachycardia (VT) episodes in 130 consecutive patients (mean age at implantation: 63.8 ± 14.9 years; sex, 95 male and 35 female) for whom implantable cardioverter defibrillator or cardiac resynchronization therapy defibrillator equipped with iATP algorithm was implanted, we investigated the efficacy and safety of iATP for VT refractory to conventional burst pacing. Eight patients had a total of 17 episodes of iATP therapy after failed conventional burst pacing within 11.2 ± 6.6 months of follow‐up. Eleven VT episodes (64.7%) in seven patients (87.5%) were successfully terminated by iATP, and only one patient (12.5%) experienced VT acceleration. iATP might be useful for VTs refractory to conventional burst pacing with a low risk of VT acceleration.
               