Many studies of cardiovascular function require a realistic representation of vascular geometry. Corrosion casting has been used to acquire such geometries for many decades. However, the fidelity with which this… Click to show full abstract
Many studies of cardiovascular function require a realistic representation of vascular geometry. Corrosion casting has been used to acquire such geometries for many decades. However, the fidelity with which this method reproduces vascular anatomy has not been completely determined. Here we report on the non‐linear shrinkage characteristics and exothermic properties of Batson's #17, a widely used casting resin, in model systems and in aortas of rats and rabbits. The setting process was captured using high‐resolution photography. Shrinkage ranged from 3.4 ± 1.5% of the diameter in 1 ml plastic syringes (inner diameter 4.8 mm) to 19.6 ± 5.6% in the aorta of rats (diameter 1.5–2.6 mm). In addition, aortic curvature and branching angles changed during setting. These effects should be determined and corrected in studies of vascular geometry where high accuracy is required.
               
Click one of the above tabs to view related content.