BACKGROUND Active pleural suction devices may be required for continuous drainage of pleural fluid or air. Such devices may not always be available or economical to stock at all veterinary… Click to show full abstract
BACKGROUND Active pleural suction devices may be required for continuous drainage of pleural fluid or air. Such devices may not always be available or economical to stock at all veterinary hospitals. Three designs of pleural drainage systems adapted from standard suction canisters are proposed. KEY FINDINGS The designs were constructed from readily available materials from the veterinary hospital and consist of (1) a 1-bottle system intended for use with pneumothoraces, (2) a 2-bottle system, which can be used for draining air and fluid, and (3) an abbreviated 2-bottle system. The suction pressure was tested using a water manometer for each design. The suction pressure was consistently accurate with designs 1 and 2, with mild fluctuations, intermittently increasing suction pressure in design 1 due to bubbling through the air vent. Design 3 demonstrated a greater degree of pressure fluctuations, which was attributed to having the least water volume, reaching suction pressures up to 1 cm H2 O above what was set. SIGNIFICANCE The improvised devices appear effective, allow accurate delivery of a set suction pressure, and can be rapidly constructed from inexpensive and readily available equipment.
               