Ultrafast ultrasound imaging using plane or diverging waves, not focused beams, has enabled quantitative assessment of biological tissue elasticity, kinematics, and hemodynamics beyond anatomical information in the past decade. However,… Click to show full abstract
Ultrafast ultrasound imaging using plane or diverging waves, not focused beams, has enabled quantitative assessment of biological tissue elasticity, kinematics, and hemodynamics beyond anatomical information in the past decade. However, its sonographic signal-to-noise ratio (SNR) and penetration depth are limited by insufficient energy delivery under safety limits. We hereby propose a novel coding scheme and apply it to ultrafast ultrasound imaging to increase the SNR without compromising the spatial resolution and frame rate. In our coded ultrafast ultrasound imaging scheme, each transmit is a long pulse containing N (N = 2k, k = 0, 1, 2,…) waves with short time intervals and polarity coefficients of + 1 or −1, instead of the conventional short pulse with single wave. In reception, a linear decoding scheme comprised of addition, subtraction and delay operations is devised to recover N times higher intensity backscattered signals to gain SNR of 10·log10(N). Experimental results acquired by the Verasonics ...
               