A generalized sidelobe canceler (GSC) approach is proposed for medical ultrasound imaging. The approach uses a set of adaptive weights instead of traditional non-adaptive weights, thus suppressing the interference and… Click to show full abstract
A generalized sidelobe canceler (GSC) approach is proposed for medical ultrasound imaging. The approach uses a set of adaptive weights instead of traditional non-adaptive weights, thus suppressing the interference and noise signal of echo data. In order to verify the validity of the proposed approach, Field II is applied to obtain the echo data of synthetic aperture (SA) for 13 scattering points and circular cysts. The performance of GSC is compared with SA using boxcar weights and Hamming weights, and is quantified by the full width at half maximum (FWHM) and peak signal-to-noise ratio (PSNR). Imaging of scattering point utilizing SA, SA (hamming), GSC provides FWHMs of 1.13411, 1.68910, 0.36195 mm and PSNRs of 60.65, 57.51, 66.72 dB, respectively. The simulation results of circular cyst also show that GSC can perform better lateral resolution than non-adaptive beamformers. Finally, an experiment is conducted on the basis of actual echo data of an ultrasound system, the imaging result after SA, SA (hamming), GSC provides PWHMs of 2.55778, 3.66776, 1.01346 mm at z = 75.6 mm, and 2.65430, 3.76428, 1.27889 mm at z = 77.3 mm, respectively.
               