A novel concept for a superconducting, fixed-field bending section is presented for use in a proton therapy gantry. The large momentum acceptance of this design allows for treatment over the… Click to show full abstract
A novel concept for a superconducting, fixed-field bending section is presented for use in a proton therapy gantry. The large momentum acceptance of this design allows for treatment over the full proton energy range of 70-220 MeV with fixed field in the superconducting magnets, eliminating the technical risks associated with fast-field ramping to match beam energy changes during treatment. A combined study of beam dynamics and magnet design is shown for such a system in which a simple magnet geometry with straight Nb-Ti racetrack coils is used to produce the desired fields. Particle tracking through this design is compared with clinical requirements for beam spot shape and size at isocenter over the full range of proton energy.
               
Click one of the above tabs to view related content.