The American Academy of Pediatrics recommends screening newborns ≥35 weeks’ gestation with total serum bilirubin (TSB) or transcutaneous bilirubin (TcB) to detect hyperbilirubinemia. Retrospective studies show TcB measurements strongly correlate… Click to show full abstract
The American Academy of Pediatrics recommends screening newborns ≥35 weeks’ gestation with total serum bilirubin (TSB) or transcutaneous bilirubin (TcB) to detect hyperbilirubinemia. Retrospective studies show TcB measurements strongly correlate with TSB; however, few prospective trials document this relationship. Furthermore, Dräger’s newest TcB instrument, JM-105, remains unstudied in the United States. We measure TcB on foreheads and sternums of newborns using JM-105 and Bilichek devices within 30 minutes of TSB measurement. We find best overall TcB/TSB correlation with JM-105 on the sternum (mean TcB-TSB difference: −0.21 ± 1.15 mg/dL). Correlations between paired measurements for TcB on the sternum using JM-105 were 0.93 for all TSB levels (n = 178), 0.82 for TSB > 10 (n = 19), 0.69 for TSB > 12 (n = 11), and 0.52 for TSB > 15 (n = 6). TcB accuracy via JM-105 on the sternum significantly differed among races (P < .001). For 5% of paired measurements, TcB with JM-105 on the sternum underestimated TSB by ≥2 mg/dL, and for <1% by ≥3 mg/dL.
               