The false positive rates for detecting keratoconus and potential ectatic corneal conditions when evaluating astigmatic eyes with Scheimpflug Technology

Photo from wikipedia

Purpose To quantify the false positive rates for keratoconus (KC) and potential ectatic corneal conditions in highly astigmatism eyes when using published parameters/indices obtained from the Pentacam and Galilei units.… Click to show full abstract

Purpose To quantify the false positive rates for keratoconus (KC) and potential ectatic corneal conditions in highly astigmatism eyes when using published parameters/indices obtained from the Pentacam and Galilei units. Setting Oftalmosalud Instituto de Ojos, Lima, Peru. Design Prospective cohort study. Methods 67 consecutive eyes with corneal astigmatism > 1.5 D, with a minimum follow ups of 36 months after an uneventful LASIK procedure were included. Indices for KC and other potential ectatic corneal conditions (subclinical KC, forme fruste KC, suspect KC) were obtained using the Pentacam and Galilei Scheimpflug cameras. Main outcome measures The false positive rates for KC and potential ectatic corneal conditions were measured. Cut off values provided by previous studies and company-based parameters were used to assess the rate of false positivity. Results The range of false positive rates for a KC diagnosis depending on the lowest and highest cutoff values were: index of height decentration (61% - 1%), index of surface variance (76% - 0%), Posterior elevation (55% - 0%), maximum Ambrosio Relational thickness (100% - 13%), Belin Ambrosio enhanced ectasia display total deviation value (100% - 4%), Average pachymetric progression index (69% - 3%), Pachymetry at the thinnest point (58% - 1%), CSI Center Surround Index (100%), Differential sector index (51%). Conclusion The false positive rates for KC and ectatic corneal conditions vary dramatically depending on the cut-off values used. Some indexes used for diagnosis of potential ectatic corneal conditions are inaccurate in normal, highly astigmatic eyes.

Keywords: potential ectatic; corneal conditions; ectatic corneal; false positive; positive rates

Journal Title: European Journal of Ophthalmology
Year Published: 2022

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.