Purpose To show an alternative surgical technique for the introduction of the intravitreal fluocinolone acetonide (FAc) implant (Iluvien®) into the vitreous cavity using a 23-gauge (G) trocar if it is… Click to show full abstract
Purpose To show an alternative surgical technique for the introduction of the intravitreal fluocinolone acetonide (FAc) implant (Iluvien®) into the vitreous cavity using a 23-gauge (G) trocar if it is retained during its implantation in the subconjunctival space. Methods We describe the surgical procedure performed to solve the complication: The FAc implant was extracted from the subconjunctival space using flat retinal forceps. A 23-G trocar was inserted 3,5 mm to the limbus. The same flat retinal forceps were used to take the FAc implant and introduce it into the vitreous cavity using a 23-G trocar. Results The patient's best corrected visual acuity (BCVA) (Snellen) improved from 20/200 to 20/63 and the central macular thickness (CMT) was reduced from 610 microns (µm) to 215 µm after one week of the FAc implantation. He remained stable after 3 months of follow-up, with a BCVA of 20/63 and a CMT of 191 µm. His intraocular pressure (IOP) remained stable and the integrity of the implant was checked by indirect ophthalmoscopy. Conclusion The introduction of the intravitreal FAc implant using a 23-gauge trocar constitutes a valid alternative if it is retained during its implantation in the subconjunctival space. The functionality of the implant remained intact in our patient.
               