LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

Metal artifact reduction by monoenergetic extrapolation of dual-energy CT in patients with metallic implants

Photo by sharonmccutcheon from unsplash

Purpose: The objective of this study is to assess artifact reduction and image quality using dual-energy computed tomography (DECT) and metal artifact reduction techniques in patients with metallic implants. Methods:… Click to show full abstract

Purpose: The objective of this study is to assess artifact reduction and image quality using dual-energy computed tomography (DECT) and metal artifact reduction techniques in patients with metallic implants. Methods: Forty patients with metallic implants, who had targeted CT performed by DECT during March to September 2018, were prospectively recruited. Post-processing with monoenergetic extrapolation at 70 and 150 keV was performed. Forty matched controls with metallic implants with single-energy CT (SECT) performed were selected. Attenuation value, noise, and signal-to-noise ratio (SNR) at the site of maximal artifact were measured at muscle and fat areas. Image quality of three sets of images (70 keV, 150 keV, and SECT) was assessed by two independent reviewers using a 5-point Likert-type scale. Statistical analysis of measured values, Likert-type scales, and radiation doses (volume CT dose index (CTDIvol)) of DECT and SECT were performed with Mann–Whitney U test. Results: As compared to SECT, high keV reconstruction of DECT show (1) significantly higher values within muscle and fat surrounding the implant (DECT vs. SECT—muscle: −96 Hounsfield units (HU) vs. −405 HU, fat: −115 HU vs. −301 HU; p < 0.001), (2) significantly lower mean image noise (75 HU vs. 129 HU; p = 0.02), and (3) higher SNR (−0.8 vs. −4.3; p < 0.001). In addition, image quality of high keV reconstruction was rated superior to the other two groups on Likert-type scales (p < 0.001). The mean radiation doses (CTDIvol) were comparable between DECT and SECT (14.2 mGy vs. 19.3 mGy; p = 0.08). Conclusion: For patients with metallic implants, monoenergetic extrapolation of DECT at high keV can reduce metal artifacts, increase SNR, and improve qualitative image quality at comparable radiation dose.

Keywords: patients metallic; monoenergetic extrapolation; dect; artifact reduction; metallic implants

Journal Title: Journal of Orthopaedic Surgery
Year Published: 2019

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.