Background To evaluate the effect of topical prostaglandin analogues on agreement of IOP measurements obtained by Goldmann applanation tonometry (GAT), rebound tonometry (RBT), and noncontact tonometry (NCT) in eyes with… Click to show full abstract
Background To evaluate the effect of topical prostaglandin analogues on agreement of IOP measurements obtained by Goldmann applanation tonometry (GAT), rebound tonometry (RBT), and noncontact tonometry (NCT) in eyes with primary open- angle glaucoma (POAG). Methods Intraocular pressure measurements were obtained using GAT, RBT, and NCT in patients with POAG with or without prostaglandin analogues. The agreement between each tonometry was analysed using Bland-Altman analyses in those with or without prostaglandin analogues. The effect of average IOP on IOP differences was also evaluated. Results Among a total of 86 subjects included in the study, 44 patients were using prostaglandin analogues. The difference in IOP measured by GAT and RBT was marginally greater in those with (GAT-RBT: − 0.94 ± 1.63 mmHg) prostaglandin analogues than in those without (− 0.33 ± 1.22 mmHg, P = 0.06). The difference in IOP measured by GAT and NCT was significantly greater in the prostaglandin group (GAT-NCT: 2.40 ± 2.89 mmHg) than in the group without prostaglandin analogues (0.41 ± 1.63 mmHg, P < 0.01). While there was no significant relationship between the average of all tonometries and the difference between tonometries in those without prostaglandin analogues, both RBT and NCT underestimated IOP relative to GAT at higher IOP in those using prostaglandin analogues. Conclusion Intraocular pressure measured by RBT and NCT was similar to that measured by GAT in those without prostaglandin analogues. RBT overestimated and NCT underestimated IOP compared to GAT in those using prostaglandin analogues.
               
Click one of the above tabs to view related content.