Optical heterodyne detection-based spectrometers are attractive due to their relatively simple construction and ultrahigh resolution. Here we demonstrate a proof-of-principle single-mode optical-fiber-based heterodyne spectrometer that has picometer resolution and quantum-limited… Click to show full abstract
Optical heterodyne detection-based spectrometers are attractive due to their relatively simple construction and ultrahigh resolution. Here we demonstrate a proof-of-principle single-mode optical-fiber-based heterodyne spectrometer that has picometer resolution and quantum-limited sensitivity around 1550 nm. Moreover, we report a generalized quantum limit of detecting broadband multispectral-temporal-mode light using heterodyne detection, which provides a sensitivity limit on a heterodyne detection-based optical spectrometer. We then compare this sensitivity limit to several spectrometer types and dim light sources of interest such as spontaneous parametric downconversion, Raman scattering, and spontaneous four-wave mixing. We calculate that the heterodyne spectrometer is significantly less sensitive than a single-photon detector and is unable to detect these dim light sources, except for the brightest and narrowest-bandwidth examples.
               