LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

Low-cost hand-held probe for depth-resolved low-coherence interferometry.

Photo from wikipedia

We report on the development of a low-cost hand-held low-coherence interferometric imaging system based on the principle of linear optical coherence tomography (Linear OCT), a technique which was first proposed… Click to show full abstract

We report on the development of a low-cost hand-held low-coherence interferometric imaging system based on the principle of linear optical coherence tomography (Linear OCT), a technique which was first proposed in the early 2000s as a simpler alternative to the conventional time-domain and Fourier-domain OCT. A bench-top implementation of the proposed technique is first presented and validated. The axial resolution, SNR, and sensitivity roll-of of the system was estimated to be 5.2 μm and 80 dB, and 3.7 dB over a depth of 0.15 mm, respectively. After validating the bench-top system, two hand-held probe implementations for contact-based imaging and in vivo human tympanic membrane imaging are presented. The performance of the proposed system was compared with a research-grade state-of-the-art Fourier-domain low coherence interferometry (LCI) system by imaging several biological and non-biological samples. The results of this study suggest that the proposed system might be a suitable choice for applications where imaging depth and SNR can be traded for lower cost and simpler optical design.

Keywords: system; coherence; hand held; low coherence; cost

Journal Title: Biomedical optics express
Year Published: 2017

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.