PURPOSE To compare the accuracy of toric intraocular lens (IOL) calculations on three modern toric calculators. MATERIAL AND METHODS The study comprised 35 eyes of 35 patients who underwent phacoemulsification with toric IOL implantation (EnVista Toric). Residual postoperative refractive astigmatism was calculated on three calculators: EnVista Toric Calculator, ASSORT Toric IOL Calculator and the Kane formula. Prediction error for each calculator was determined using vector analysis. RESULTS The mean absolute deviation of predicted postoperative refractive astigmatism over actual astigmatism in diopters was distributed in the following way: 0.82±0.58, 0.70±0.67 and 0.72±0.76 using EnVista Toric Calculator, ASSORT Toric IOL Calculator and the Kane formula, respectively. Centroid prediction error was 0.08 (EnVista Toric Calculator), 0.06 (ASSORT) and 0.10 (Kane formula). There was a significantly smaller deviation using ASSORT and the Kane formula compared to the online calculator (p<0.05). CONCLUSIONS Toric calculators ASSORT Toric IOL Calculator and the Kane formula showed higher accuracy of toric IOL calculation than EnVista Toric Calculator.
               