LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

A precise and rapid early pregnancy test: Development of a novel and fully automated electrochemical point-of-care biosensor for human urine samples.

Photo from wikipedia

Human chorionic gonadotropin (hCG), a glycoprotein hormone secreted from the placenta, is an important biomarker for pregnancy. In this study, we designed a precise, rapid and fully automatic device with… Click to show full abstract

Human chorionic gonadotropin (hCG), a glycoprotein hormone secreted from the placenta, is an important biomarker for pregnancy. In this study, we designed a precise, rapid and fully automatic device with an electrochemical point-of-care biosensor capable of quantitative hCG detection from human urine samples for early pregnancy detection. Gold and Ag/AgCl electrodes, whose structure with optimum isopotential region and current density, were simulated using COMSOL Multiphysics® software and custom-made from Flex Medical. The sensing surface was fabricated with DSP self-assembled monolayers (SAMs) and covalently immobilized anti-hCG-beta antibody. The detection method involved a sandwich assay using anti-hCG alpha-HRP. Based on an automated agitation design implemented in our device, the surface reaction rate is significantly improved comparing to routinely performed sandwich assays, and therefore a rapid detection of very low concentration can be achieved. Electrochemical impedance spectroscopy (EIS) and chronoamperometry (CA) measurements were used to characterize the immobilization of the antibodies and to determine the sensor activities respectively. The sensors displayed a limit of detection (LOD) of 2.17 mIU/ml within established clinical hCG levels for early detection of pregnancy. They responded very well to hCG, but not to luteinizing hormone (LH), which has a high degree of cross-reactivity with hCG. The results showed that the immunosensor has high specificity, good reproducibility, and long-term stability for the detection of hCG in urine samples.

Keywords: precise rapid; hcg; urine samples; electrochemical point; detection; pregnancy

Journal Title: Talanta
Year Published: 2022

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.