3D cine phase-contrast (4D flow) MRI is a sequence with great potential for non-invasive time-resolved 3D flowmetry at arbitrary vessel sections in various blood vessels. However, it is not widely… Click to show full abstract
3D cine phase-contrast (4D flow) MRI is a sequence with great potential for non-invasive time-resolved 3D flowmetry at arbitrary vessel sections in various blood vessels. However, it is not widely known that the flowmetry with 4D flow MRI is vulnerable to pulsatile and non-uniform flow. Due to the limited spatial and temporal resolutions, averaging within the 3D voxel is occurring during the flowmetry. A simple solution is to avoid setting the measurement plane in the area where non-uniform flow is dominant, which is possible with an aid of streamline depictions generated by computational fluid dynamics (CFD) or 4D flow MRI data. Unlike 4D flow MRI, flowmetry in CFD simulation can use higher spatial and temporal resolution depending on computer performance; therefore, it is robust to fluctuating non-uniform flow. However, the performance of CFD simulations might be limited due to inlet conditions with low temporal resolution. Difficulty applying complex blood flow such as reflection flow from periphery may also limit accurate simulation. Caution should be taken when comparing the result of CFD simulation to that of 4D flow measurement.
               