BACKGROUND Recent advances in exoskeleton technology has made lower extremity powered exoskeletons (LEPE) a viable treatment tool to restore upright walking mobility to persons with spinal cord injury (SCI). OBJECTIVE… Click to show full abstract
BACKGROUND Recent advances in exoskeleton technology has made lower extremity powered exoskeletons (LEPE) a viable treatment tool to restore upright walking mobility to persons with spinal cord injury (SCI). OBJECTIVE Evaluate ARKE exoskeleton training within a rehabilitation centre environment. METHODS Case studies are presented for two male participants, age 41 and 30, motor complete SCI at T6 (N01) and T12 (N02), respectively, as they progress from new LEPE users to independent walking. The ARKE 2.0 LEPE (Bionik Laboratories Inc., Toronto, Canada) was used for all training (hip and knee powered, forearm crutches, control tablet). Data were collected on session times, activity metrics from ARKE system logs, and qualitative questionnaire feedback. RESULTS AND CONCLUSION N01 required 18, 30-minute training sessions to achieve independent walking. N01 walked independently within the 12 session target. Foot strikes were frequently before the end of the programmed swing phase, which were handled by the ARKE control system. Subjective ratings of LEPE learning, comfort, pain, fatigue, and overall experience were high for sitting-standing and moderate to high for walking. This reflected the complexity of learning to safely walk. Qualitative feedback supported the continuation of LEPE use in rehabilitation settings based on end-user desire for upright mobility.
               