The detection and quantification of hepatitis B virus (HBV) DNA plays an important role in diagnosing and monitoring HBV infection as well as in assessing the therapeutic response. We compared… Click to show full abstract
The detection and quantification of hepatitis B virus (HBV) DNA plays an important role in diagnosing and monitoring HBV infection as well as in assessing the therapeutic response. We compared the analytical performance of a random access, fully automated HBV assay—DxN VERIS Molecular Diagnostics System (Beckman Coulter, Brea, CA, USA)—with that of Abbott RealTime HBV assay (Abbott Laboratories, Des Plaines, IL, USA). The between-day precision of the VERIS assay ranged from 0.92% (mean 4.68 log IU/mL) to 4.15% (mean 2.09 log IU/mL) for pooled sera from HBV patients. HBV DNA levels measured by the VERIS HBV assay correlated with the calculated HBV DNA levels (r2=0.9994; P<0.0001). The lower limit of quantification was estimated as 8.76 IU/mL (Probit analysis, 95% confidence interval: 7.32–12.00 IU/mL). Passing-Bablok regression analysis showed good concordance between the VERIS and RealTime assays for 187 chronic HBV samples (y=−0.2397+0.9712x; r=0.981), as well as for 20 drug-resistant HBV genotype C positive samples (y=−0.5415+0.9954x; r=0.961). The VERIS assay demonstrated performance similar to the RealTime assay and is suitable for high-throughput HBV DNA monitoring in large hospital laboratories.
               