Comparison of the International Normalized Ratio Between a Point-of-Care Test and a Conventional Laboratory Test: the Latter Performs Better in Assessing Warfarin-induced Changes in Coagulation Factors

Photo from wikipedia

Background Point-of-care testing (POCT) coagulometers are increasingly used for monitoring warfarin therapy. However, in high international normalized ratio (INR) ranges, significant discrepancy in the INR between POCT and conventional laboratory… Click to show full abstract

Background Point-of-care testing (POCT) coagulometers are increasingly used for monitoring warfarin therapy. However, in high international normalized ratio (INR) ranges, significant discrepancy in the INR between POCT and conventional laboratory tests occurs. We compared the INR of POCT (CoaguChek XS Plus; Roche Diagnostics, Mannheim, Germany) with that of a conventional laboratory test (ACL TOP 750; Instrumentation Laboratory SpA, Milan, Italy) and explored possible reasons for discrepancy. Methods Paired POCT and conventional laboratory test INRs were analyzed in 400 samples from 126 patients undergoing warfarin therapy after cardiac surgery. Coagulation factor and thrombin generation tests were compared using the Mann–Whitney U test. Correlations between coagulation factors and INRs were determined using Pearson correlation coefficients. Results The mean difference in the INR between the tests increased at high INR ranges. Endogenous thrombin potential levels were decreased at INR <2.0 for CoaguChek XS Plus and 2.0< INR <3.0 for ACL TOP 750 compared with those at INR <2.0 for both tests, indicating a better performance of ACL TOP 750 in assessing thrombin changes. The correlation coefficients of coagulation factors were stronger for ACL TOP 750 INR than for CoaguChek XS Plus INR. Vitamin K-dependent coagulation factors were found to contribute to the INR discrepancy. Conclusions Decreases in vitamin K-dependent coagulation and anticoagulation factors can explain the significant discrepancy between the two tests in high INR ranges. Since conventional laboratory test INR values are more reliable than POCT INR values, a confirmatory conventional laboratory test is required for high INR ranges.

Keywords: inr; coagulation factors; conventional laboratory; laboratory test

Journal Title: Annals of Laboratory Medicine
Year Published: 2023

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.