Conventional methods that have been developed to immobilize the mouth and tongue for radiotherapy (RT) in head and neck cancer (HNC) treatment have been unsatisfactory. We, therefore, developed three-dimensional (3D),… Click to show full abstract
Conventional methods that have been developed to immobilize the mouth and tongue for radiotherapy (RT) in head and neck cancer (HNC) treatment have been unsatisfactory. We, therefore, developed three-dimensional (3D), customizable, silicone bite blocks and examined their clinical feasibility. For HNC patients, before RT, the 3D printed bite blocks were fabricated based on primary computed tomography (CT) simulation images. The placement of the 3D bite blocks was followed by a secondary CT simulation before RT planning was finalized. Dosimetric parameters and positioning verification achieved with the propose bite blocks were compared with conventional universal oral corks. The 3D printed bite blocks were conformal to the occlusal surface, ensuring immobilization of the tongue without eliciting a gag reflex, and an elastic and firm texture that supports opening of the mouth, with a smooth surface with tolerable intraoral tactility. The dosimetry of patients using the proposed bite blocks showed better coverage of the planning target volume and surface of a tumour bed along with reduction in normal tissue doses. Good concordance of positioning by 3D printed bite blocks during the RT course was verified. The 3D printed bite blocks with silicone might be a customizable, safe, and practical advanced technology in RT for HNC.
               