This study aims to investigate the qualitative and quantitative image quality of low-dose high-resolution (LD-HR) lung CT scans acquired with the first clinical approved photon counting CT (PCCT) scanner. Furthermore,… Click to show full abstract
This study aims to investigate the qualitative and quantitative image quality of low-dose high-resolution (LD-HR) lung CT scans acquired with the first clinical approved photon counting CT (PCCT) scanner. Furthermore, the radiation dose used by the PCCT is compared to a conventional CT scanner with an energy-integrating detector system (EID-CT). Twenty-nine patients who underwent a LD-HR chest CT scan with dual-source PCCT and had previously undergone a LD-HR chest CT with a standard EID-CT scanner were retrospectively included in this study. Images of the whole lung as well as enlarged image sections displaying a specific finding (lesion) were evaluated in terms of overall image quality, image sharpness and image noise by three senior radiologists using a 5-point Likert scale. The PCCT images were reconstructed with and without a quantum iterative reconstruction algorithm (PCCT QIR+/−). Noise and signal-to-noise (SNR) were measured and the effective radiation dose was calculated. Overall, image quality and image sharpness were rated best in PCCT (QIR+) images. A significant difference was seen particularly in image sections of PCCT (QIR+) images compared to EID-CT images (p < 0.005). Image noise of PCCT (QIR+) images was significantly lower compared to EID-CT images in image sections (p = 0.005). In contrast, noise was lowest on EID-CT images (p < 0.001). The PCCT used significantly less radiation dose compared to the EID-CT (p < 0.001). In conclusion, LD-HR PCCT scans of the lung provide better image quality while using significantly less radiation dose compared to EID-CT scans.
               
Click one of the above tabs to view related content.