The purpose of this research was to enable component separation based on simple control of the flow rate. We investigated a method that eliminated the need for a centrifuge and… Click to show full abstract
The purpose of this research was to enable component separation based on simple control of the flow rate. We investigated a method that eliminated the need for a centrifuge and enabled easy component separation on the spot without using a battery. Specifically, we adopted an approach that uses microfluidic devices, which are inexpensive and highly portable, and devised the channel within the fluidic device. The proposed design was a simple series of connection chambers of the same shape, connected via interconnecting channels. In this study, polystyrene particles with different sizes were used, and their behavior was evaluated by experimentally observing the flow in the chamber using a high-speed camera. It was found that the objects with larger particle diameters required more time to pass, whereas the objects with smaller particle diameters flowed in a short time; this implied that the particles with a smaller size could be extracted more rapidly from the outlet. By plotting the trajectories of the particles for each unit of time, the passing speed of the objects with large particle diameters was confirmed to be particularly low. It was also possible to trap the particles within the chamber if the flow rate was below a specific threshold. By applying this property to blood, for instance, we expected plasma components and red blood cells to be extracted first.
               