Herein, we report a new automated flow method based on zone fluidics for the simultaneous determination of homocysteine and homocysteine thiolactone using fluorimetric detection (λext = 370 nm/λem = 480… Click to show full abstract
Herein, we report a new automated flow method based on zone fluidics for the simultaneous determination of homocysteine and homocysteine thiolactone using fluorimetric detection (λext = 370 nm/λem = 480 nm). Homocysteine thiolactone is hydrolyzed on-line in alkaline medium (1 mol L−1 NaOH) to yield homocysteine, followed by reaction with o-phthalaldehyde in a single step. Derivatization is rapid without the need of elevated temperatures and stopped-flow steps, while specificity is achieved through a unique reaction mechanism in the absence of nucleophilic compounds. Mixtures of the analytes can be analyzed quantitatively after specific separation with fluorosurfactant-capped gold nanoparticles that are selectively aggregated by homocysteine, leaving the thiolactone analogue in solution. As low as 100 nmol L−1 of the analyte(s) can be quantified in aqueous solutions, while concentrations > 2 μmol L−1 can be analyzed in artificial and real urine matrix following 20-fold dilution. The percent recoveries ranged between 87 and 119%.
               
Click one of the above tabs to view related content.