We develop an electrochemical sensor for the determination of bromhexine hydrochloride (BHC), a widely use mucolytic drug. The sensor is prepared by electrodeposition of cobalt oxides (CoOx) on a glassy… Click to show full abstract
We develop an electrochemical sensor for the determination of bromhexine hydrochloride (BHC), a widely use mucolytic drug. The sensor is prepared by electrodeposition of cobalt oxides (CoOx) on a glassy carbon electrode modified with carboxylated single-walled carbon nanotubes (SWCNT). A synergistic effect between CoOx and SWCNT is observed, leading to a significant improvement in the BHC electrooxidation current. Based on cyclic voltammetry studies at varying scan rates, we conclude that the electrochemical oxidation of BHC is under mixed diffusion–adsorption control. The proposed sensor allows the amperometric determination of BHC in a linear range of 10–500 µM with a low applied voltage of 0.75 V. The designed sensor provides reproducible measurements, is not affected by common interfering substances, and shows excellent performance for the analysis of BHC in pharmaceutical preparations.
               
Click one of the above tabs to view related content.