LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

Evaluation of the measurement precision and accuracy in the dental CAD/CAM system.

Photo from wikipedia

The purpose of this study is to investigate the precision and trueness of the occlusal contact record obtained with a dental CAD/CAM device (ARCTICA). Sandblasted spherical steel balls with diameters… Click to show full abstract

The purpose of this study is to investigate the precision and trueness of the occlusal contact record obtained with a dental CAD/CAM device (ARCTICA). Sandblasted spherical steel balls with diameters of 10-20 mm were measured using ARCTICA and a three-dimensional measurement device (FN503). The radius of each steel ball was calculated from the measured value and the difference between the measured and nominal values was obtained. Upper/lower dental arch casts were measured and processed for occlusal contact image creation using both devices. The graphical images of the occlusal contacts at the intercuspal position obtained from both systems and an occlusal analysis device (BE-1) were compared with the occlusal contact area. Excellent correlation was observed between the measurement results of ARCTICA and FN503 (p<0.001, R2: 0.99). The occlusal contact areas were 186.0 mm2 (ARCTICA), 192.8 mm2 (FN503), and 196.1 mm2 (BE-1). This study showed that ARCTICA is capable of conducting highly accurate measurements and generating a graphical image of occlusal contacts with good reliability.

Keywords: cad cam; precision; occlusal contact; dental cad

Journal Title: Dental materials journal
Year Published: 2020

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.