LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness.

Photo from wikipedia

Purpose To compare readings of intraocular pressure (IOP) taken with the Goldmann applanation tonometer (GAT), the non-contact tonometer (NCT), and the rebound tonometer (RBT), and to compare their correlation with… Click to show full abstract

Purpose To compare readings of intraocular pressure (IOP) taken with the Goldmann applanation tonometer (GAT), the non-contact tonometer (NCT), and the rebound tonometer (RBT), and to compare their correlation with central corneal thickness (CCT). Methods This was a prospective, cross-sectional, observational study to which patients above 18 years of age were enrolled. A total of 400 eyes of 200 non-glaucomatous patients underwent IOP recordings using the GAT, NCT, and RBT, and CCT was also noted. Informed consent of the patients was taken. The IOP readings taken via the three methods were compared and correlated with CCT. Paired t test was used to compare the two devices. Simple and multivariate linear regression analyses were used to study the relationship between factors. A P value less than 0.05 was considered significant. Correlation was determined using the Pearson correlation coefficient, and a Bland-Altman graph was plotted. Results Mean IOP measured by the NCT was 15.65 ± 2.80 mmHg, by the RBT was 14.23 ± 3.05 mmHg, and by the GAT was 14.69 ± 2.97 mmHg. The mean CCT was 510.61 ± 33.83 microns. The difference between mean IOP recorded by the NCT and that by the RBT was 1.41 ± 2.39 mmHg, between the NCT and GAT was 0.95 ± 2.03 mmHg, and between the GAT and RBT was 0.45 ± 2.22 mmHg. The difference between the IOP values was statistically significant (P < 0.005). All tonometers showed a statistically significant correlation with CCT, but it was observed that the NCT had a stronger correlation (0.4037). Conclusion The IOP readings taken by all the three methods were comparable; however, RBT values were closer to GAT values. CCT did influence the IOP values, and this should be kept in mind while evaluating.

Keywords: gat; mmhg; tonometer; iop; rbt; correlation

Journal Title: Indian journal of ophthalmology
Year Published: 2023

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.