LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

A sensitive, robust and high-throughput isotope dilution LC-MS/MS method for quantifying three folate forms in serum.

This study aimed to establish an isotope dilution LC-MS/MS method for the determination of folic acid, 5-formyltetrahydrofolate and 5-methyltetrahydrofolate in human serum. This method was then used to quantify these… Click to show full abstract

This study aimed to establish an isotope dilution LC-MS/MS method for the determination of folic acid, 5-formyltetrahydrofolate and 5-methyltetrahydrofolate in human serum. This method was then used to quantify these three folate forms in the healthy adult population and supplement users. A stable 96-well solid-phase extraction system was used to prepare serum samples. The highly sensitive method was established using a Shimadzu LCMS-8060NX. The linearity was good in the range of 0.1-10 nmol/l for folic acid and 5-formyltetrahydrofolate and 1.0-100 nmol/l for 5-methyltetrahydrofolate. The accuracy and precision were good. The method was sensitive, robust and high-throughput and could be used for the routine clinical monitoring of these three folate forms in the Chinese population.

Keywords: isotope dilution; dilution method; folate forms; sensitive robust; method; three folate

Journal Title: Bioanalysis
Year Published: 2023

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.