Purpose To report an implementation method and the results of independent brachytherapy dose verification software (DVS). Material and methods The DVS was developed based on Visual C++ and adopted a… Click to show full abstract
Purpose To report an implementation method and the results of independent brachytherapy dose verification software (DVS). Material and methods The DVS was developed based on Visual C++ and adopted a modular structure design. The DICOM RT files exported from a treatment planning system (TPS) were automatically loaded into the DVS. The DVS used the TG-43 formalism for dose calculation. A total of 15 cervical cancer patients who underwent brachytherapy were retrospectively selected to test the DVS. Dosimetric parameters and γ analysis (0.1 cm, 5%) were used to evaluate the dose differences between the DVS and the TPS. Results Compared with the TPS dose, the γ pass rates of the dose calculated by the DVS were higher than 98%. For the clinical target volume (CTV), the dosimetric differences were less than 0.63% for D90% and D100%. For the bladder, rectum, and sigmoid, the agreement of D0.1cc, D1cc, and D2cc were within a 0.78% level. Conclusions With minimal human-computer interactions, the DVS can verify the accuracy of doses calculated by the TPS.
               