Abstract. Atmospheric Hg measurements are commonly carried out using Tekran® Instruments Corporation's model 2537 Hg vapor analyzers, which employ gold amalgamation preconcentration sampling and detection by thermal desorption (TD) and… Click to show full abstract
Abstract. Atmospheric Hg measurements are commonly carried out using Tekran® Instruments Corporation's model 2537 Hg vapor analyzers, which employ gold amalgamation preconcentration sampling and detection by thermal desorption (TD) and atomic fluorescence spectrometry (AFS). A generally overlooked and poorly characterized source of analytical uncertainty in those measurements is the method by which the raw Hg atomic fluorescence (AF) signal is processed. Here I describe new software-based methods for processing the raw signal from the Tekran® 2537 instruments, and I evaluate the performances of those methods together with the standard Tekran® internal signal processing method. For test datasets from two Tekran® instruments (one 2537A and one 2537B), I estimate that signal processing uncertainties in Hg loadings determined with the Tekran® method are within ±[1 % + 1.2 pg] and ±[6 % + 0.21 pg], respectively. I demonstrate that the Tekran® method can produce significant low biases (≥ 5 %) not only at low Hg sample loadings (
               