BACKGROUND Monopolar electrocautery is an important tool for harvesting the pedicular internal thoracic artery (ITA) in cardiac surgery. The different power outputs of cautery may affect graft integrity and long-term… Click to show full abstract
BACKGROUND Monopolar electrocautery is an important tool for harvesting the pedicular internal thoracic artery (ITA) in cardiac surgery. The different power outputs of cautery may affect graft integrity and long-term patency. This study aimed to identify the optimal threshold of electrocautery power for ITA harvest. METHODS This prospective study included 30 patients who underwent elective coronary artery bypass surgery at the Medipol Mega University Hospital. The ITA was harvested by monopolar electrocautery after a median sternotomy. The output of cautery was adjusted at 20 W in group A and 40 W in group B. Three to 4 cm of a distal ITA sample from each patient was examined under a light microscope by two independent pathologists. RESULTS The ITA harvest time was longer in group A (21.2 ± 7.5 vs 10.3 ± 8.1 min, p < 0.001) than in group B. ITA free flow was similar in the two groups (43.6 ± 48.7 vs 51.7 ± 45.0 Ml/min, p = 0.762). Mild to moderate injury in the endothelial and sub-endothelial sample was more frequent in the low-cautery group (p = 0.0037). CONCLUSIONS ITA endothelial integrity was found to be better preserved with 40W electrocautery. Moreover, 20W of monopolar electrocautery may not be safe in pedicular ITA harvesting.
               
Click one of the above tabs to view related content.